Intraoperative retinal light damage reflected in electrophysiologic data

1991 
In a series of 30 unilaterally pseudophakic patients, electroretinograms and electrooculograms were recorded 6 months postoperatively. The unoperated on fellow eyes served as controls High intraoperative retinal light exposure (3.4–7.3 mW/cm2, Zeiss OPMI 6 operating microscope) caused a substantial reduction of electrophysiologic potentials. Light protection prevented deterioration of electroretinogram and electro-oculogram potentials; reducing the bulb voltage, tilting the axis of illumination, filtering short wavelengths and the use of light shields resulted in 4-log-unit lower intensities (0.8–3.7 μW/cm2).