Abstract 13087: Are All Patients Equal in Response to Intra-Aortic Balloon Counterpulsation?

2015 
Background: Intra-Aortic Balloon Counterpulsation (IABC) with a 50 cc (MEGA 50™) balloon catheter is associated with an average increase of 0.5- 0.7 liters/min in cardiac output. However, some patients (pts) have a much more dramatic response. We retrospectively analyzed our single center tertiary care experience with 150 consecutive pts undergoing IABC. Methods: Chart review for demographic, procedural, and hemodynamic data was collected for 150 pts of whom 64 had both pre and 4 hour post IABC hemodynamic measurements. The responder (R) group was defined by any positive change in cardiac output (CO) and cardiac index (CI) between baseline prior to IABC and 4 hours post initiation. Non-responders (NR) were defined as those with a decline in CO or no change. Results: IABC with a 50 cc balloon was associated with a significant improvement in CO of 0.7 L/min for the overall cohort (Pre-IABC mean CO 3.9±1.4 vs post 4.6 ±1.6 L/Min, paired t-test p=.0004). There were 38 pts in the R group (60 %) and 26 in the N...
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    0
    References
    0
    Citations
    NaN
    KQI
    []