[Experimental study of safety in application of the Iris Medical IQ 810 diode laser in clinical treatment of age-related macular degeneration].

2011 
Abstract During the study of diode laser radiation effect in micropulse mode on culture cells of human retinal pigment epithelium it was revealed that the quota of dead cells was a minimum. Besides, a certain conformity between dead cells quota and parameter characteristics of laser radiation. Based on the performed experimental study it was revealed that for a work using the Iris Medical IQ 810 diode laser in the micropulse mode following parameters: duration of pulse set--300ms, duration of function--9.1%, power--750mW are safe for retinal pigment epithelium cells. Rationales of safety in application of the infrared diode laser radiation in micropulse mode in clinic for treatment of age-related macular degeneration (AMD) exampled by cell culture of human retinal pigment epithelium.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    0
    References
    0
    Citations
    NaN
    KQI
    []