Couette shearing device for the investigation of shear-induced damage of the primary hemostasis by left ventricular assist devices:

2019 
Introduction:Continuous-flow left ventricular assist devices have evolved from short-time therapy into permanent or so-called destination therapy. One complication in long-term usage is bleeding, which is presumably attributed to shear-induced interference of left ventricular assist devices with the coagulation system.Methods:The influence of dynamic shear stresses on primary hemostasis by single or multiple passes through left ventricular assist devices was investigated. A novel Couette-type shearing device, especially fitted to simulate left ventricular assist devices with highly dynamic and repetitive stresses, was developed. To evaluate the clotting ability of the blood and thus the bleeding tendency, the closure time of the platelet function analyzer (PFA-100®, Dade Behring, Marburg, Germany) was used. The relationship of the PFA-100 closure time was fitted to measurement points with shear stress and exposure time as parameters.Results:76 samples of human blood collected from four different healthy d...