The Utility of iCare HOME Tonometry for Detection of Therapy-Related Intraocular Pressure Changes in Glaucoma and Ocular Hypertension.
2021
Purpose To assess whether iCare HOME rebound tonometry can detect therapy-related changes during self-monitoring of intraocular pressure (IOP). Design Prospective clinical trial. Participants A total of 43 eyes (n = 27 subjects) with open-angle glaucoma or ocular hypertension were enrolled during standard-of-care clinic visits. Participants were grouped into control eyes managed on stable therapy (n = 18 eyes) or therapy change eyes undergoing selective laser trabeculoplasty (SLT, n = 8 eyes), initiating topical therapy (n = 8 eyes), or adding a second medication to existing monotherapy (n = 9 eyes). Methods Subjects recorded IOP 4 times daily for 1 week using iCare HOME tonometry. Upon tonometer return, subjects underwent SLT or new medication start; an additional week of iCare HOME measurements was collected after 4 to 6 weeks. Control subjects recorded an additional week of measurements after 6 weeks. Measurements were grouped into 4 time periods (5–10 am , 10 am to 3 pm , 3–8 pm , 8 pm to 1 am ). Goldmann applanation tonometry (GAT) was performed at each study visit for comparison. Main Outcome Measures Detection of therapy response defined as an IOP reduction of ≥20%. Results For eyes that demonstrated a therapy response by GAT (n = 11), iCare HOME detected a therapy response in 90.9% of eyes in ≥1 time period and 45.5% of eyes in all 4 time periods. In eyes without a GAT-measured therapy response (n = 14), iCare HOME detected a response for 71.4% (n = 10) of eyes in ≥1 time period and for 7.1% of eyes (n = 1) at all 4 time periods. In treatment eyes, intraday and interday average minimum and maximum IOP, as well as interday IOP range, were significantly reduced after therapy without a significant change in intraday IOP range. Control group eyes did not demonstrate a significant change in average IOP minimum, maximum, or range between study weeks. Conclusions Home tonometry with iCare HOME reliably detects therapy-related IOP changes in patients with glaucoma and ocular hypertension. Treatment responses correlated well with in-office GAT and may detect treatment responses missed by GAT. Intraocular pressure measurements via home tonometry provide additional clinical information regarding intraday and interday IOP fluctuation beyond standard of care in office GAT measurements. The iCare HOME is a valuable tool to monitor therapeutic efficacy in patients with glaucoma.
- Correction
- Source
- Cite
- Save
- Machine Reading By IdeaReader
24
References
0
Citations
NaN
KQI