Verification of Low-Density Lipoprotein Cholesterol Levels Measured by Anion-Exchange High-Performance Liquid Chromatography in Comparison with Beta Quantification Reference Measurement Procedure.

2020 
Background A new lipoprotein testing method based on anion-exchange HPLC (AEX-HPLC) was recently established. We verified the accuracy of LDL-C levels, a primary therapeutic target for the prevention of cardiovascular disease (CVD), measured by AEX-HPLC comparing with LDL-C levels measured by beta quantification-reference measurement procedure (BQ-RMP), homogenous assays, and calculation methods. Methods We compared LDL-C levels measured by AEX-HPLC (adLDL-Ch: LDL-Ch and IDL-Ch) and BQ-RMP using blood samples from 52 volunteers. AdLDL-Ch levels were also compared with those measurements by homogeneous assays and calculation methods (Friedewald equation, Martin equation, and Sampson equation) using blood samples from 411 participants with dyslipidemia and/or type 2 diabetes. Results The precision and accuracy of adLDL-Ch were verified by BQ-RMP. The mean percentage bias [bias (%)] for LDL-C was 1.2%, and the correlation was y = 0.990x + 3.361 (r = 0.990). These results met the acceptable range of accuracy prescribed by the National Cholesterol Education Program. Additionally, adLDL-Ch levels were correlated with LDL-C levels measured by the 2 homogeneous assays (r > 0.967) and the calculation methods (r > 0.939), in serum samples from patients with hypertriglyceridemia. Conclusions AEX-HPLC is a reliable method for measuring LDL-C levels for CVD risk in daily clinical laboratory analyses.