Experimental studies of THGEM in different Ar/CO2 mixtures

2014 
In this paper, the performance of a type of domestic THGEM (THick Gaseous Electron Multiplier) working in Ar/CO2 mixtures is reported in detail. This kind of single THGEM can provide a gain range from 100 to 1000, which is very suitable for application in neutron detection. In order to study its basic characteristics as a reference for the development of a THGEM based neutron detector, the counting rate plateau, the energy resolution and the gain of the THGEM have been measured in different Ar/CO2 mixtures with a variety of electrical fields. For the Ar/CO2(90%/10%) gas mixture, a wide counting rate plateau is achieved from 720 V to 770 V, with a plateau slope of 2.4%/100 V, and an excellent energy resolution of about 22% is obtained at the 5.9 keV full energy peak of the Fe-55 X-ray source.