Comparative Accuracy of 1.5T MRI, 3T MRI, and Static Ultrasound in Diagnosis of Small Gaps in Repaired Flexor Tendons: A Cadaveric Study.

2021 
Purpose We hypothesized that magnetic resonance imaging (MRI) would more accurately diagnose small gaps ( Methods A laceration of the flexor digitorum profundus was created in 160 fresh-frozen cadaveric digits and randomized to either an intact repair (0-mm gap) or repairs using a locked 4-strand suture repair with either 4-0 Prolene, Ethibond, or and gaps of 2, 4,or 6 mm; or no suture in which 2-, 4-, or 6-mm gaps were created without a suture crossing the repair site. We performed 1.5T and 3T MRI and static US studies; gap widths were estimated by radiologists blinded to suture presence and true gap widths. Results The 1.5 and 3.0T MRI had a lower mean error than US for gap sizes 0 and 2 mm. All 3 modalities performed similarly for 4- and 6-mm gaps. Documentation of imaging artifact worsened error, and odds of seeing artifacts were 1.72 higher with MRI than with US. Suture did not worsen artifact nor impair accuracy for any of the 3 modalities. When no suture was used, all 3 modalities significantly overestimated the true gap. Conclusions MRI is most accurate for small gaps less than 4 mm. Although all modalities overestimated gap sizes in specimens with a 0-mm gap (intact tendon repair), mean overestimation ( 3 mm) repairs than the 3T. Clinical relevance Accurate diagnosis of intact repairs or small gaps (