Measurement of the liver iron concentration in transfusional iron overload by MRI R2* and by high-transition-temperature superconducting magnetic susceptometry

2019 
Abstract Purpose To compare measurement of the liver iron concentration in patients with transfusional iron overload by magnetic resonance imaging (MRI), using R2*, and by magnetic susceptometry, using a new high-transitiontemperature (high-Tc; operating at 77 K, cooled by liquid nitrogen) superconducting magnetic susceptometer. Methods In 28 patients with transfusional iron overload, 43 measurements of the liver iron concentration were made by both R2* and high-Tc magnetic susceptometry. Results Measurements of the liver iron concentration by R2* and high-Tc magnetic susceptometry were significantly correlated when comparing all patients (Pearson's r  = 0.91, p 7 mg Fe/g liver, dry weight ( r  = 0.93, p  = 0.006). In lower ranges of liver iron, no significant correlations between the two methods were found (0 to r  = 0.2, p  = 0.37; 3.2 to 7 mg Fe/g liver, dry weight: r  = 0.41; p  = 0.14). Conclusion The lack of linear correlation between R2* and magnetic susceptibility measurements of the liver iron concentration with minimal or modest iron overload may be due to the effects of fibrosis and other cellular pathology that interfere with R2* but do not appreciably alter magnetic susceptibility.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    47
    References
    3
    Citations
    NaN
    KQI
    []