Threshold monitoring of intraoperative auditory steady-state responses for ossiculoplasty surgery

2018 
AbstractObjective: The objective of this study is to establish a reliable intraoperative auditory threshold monitoring system for ossiculoplasty surgery.Design: The chirp signal of self-designed sound field earphone (SFE) was calibrated physically and psychophysically. The interaural attenuation of the SFE was tested in patients with unilateral complete deafness and contralateral normal hearing (10 patients). Self-designed SFEs were used to measure the chirp-evoked auditory steady-state responses (Chirp-ASSR) threshold of patients (14 cases and 15 ears) with conductive hearing loss after anesthesia but before surgery.Results: The response threshold of Chirp-ASSR under anesthesia displayed a strong correlation with the hearing threshold for pure tones: the Pearson coefficients at various frequencies (1, 2, and 4 kHz) were 0.56 (p = .03), 0.82 (p < .001), and 0.90 (p < .001), respectively, and the intragroup correlation coefficients were 0.70 (p = 0.02), 0.90 (p < 0.001), and 0.95 (p < 0.001), respectively....
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    13
    References
    0
    Citations
    NaN
    KQI
    []