COLD-HRM PCR versus Conventional HRM PCR to Detect the BRAF V600E Mutation: A Real Improvement?

2011 
To the Editor-in-Chief: In the recent article by Mancini et al,1 the authors demonstrated the improved sensitivity of the detection of KRAS mutations and the BRAF V600E mutation using coamplification at lower temperature PCR (COLD-PCR) in combination with high-resolution melting (HRM) instead of conventional PCR followed by HRM. They reported an eightfold higher sensitivity for KRAS mutations (from 6.2% to 0.8%) and a debatable fourfold higher sensitivity for the BRAF V600E mutation (from 12.5% to 6.2%, which would seem to be a twofold higher sensitivity), using serial dilutions of DNA from cell lines harboring the mutations (CCRF-CEM and SKMel28) and DNA from MCF-7 as wild-type control. Using conventional PCR and HRM, we2 and others, including the authors,3 have previously demonstrated detection limits for the BRAF V600E mutation of less than 5% mutated DNA in a background of wild-type DNA, which is more sensitive than the COLD-PCR assay described in their present article. In addition, our system2 proved to be applicable to FFPE material. With FFPE samples, the spread of the melting curves increases due to the lower quality of DNA; however, we were able to distinguish samples containing 5% mutated DNA in a background of wild-type DNA. We read with great interest the publications on the advantages of COLD-PCR. Indeed, we have used an extensive optimization process to try to improve the sensitivity of the BRAF assay using COLD-PCR instead of conventional PCR. We designed new primer sets that yield short amplicons to maximize the temperature effect caused by the mismatch in the heteroduplexes during COLD-PCR, taking into account single-nucleotide polymorphisms and the pseudogene described for the BRAF gene. We optimized the reaction composition and the PCR protocol. Unfortunately, we have as yet not been able to improve the sensitivity of 5% mutated DNA (SKMel28) in wild-type DNA (human mononuclear cell DNA) by using COLD-PCR, and we could not observe the expected shift in the quantification cycle values caused by the preferential amplification of the minority alleles (unpublished data). We could, however, confirm the equal melting behavior of 100% homozygously mutated SKMel28 DNA and 100% wild-type DNA in the difference plots (unpublished data). Thus, with regard to Figure 1 (right) in the study by Mancini et al,1 we do not understand the different melting behavior of curves A (MCF-7 wild-type DNA) and G (100% SKMel28) because no heteroduplexes can be formed in these reactions and the T-A substitution should not cause any alteration in the melting behavior (in consideration of the hydrogen bonds). Thus, although COLD-PCR has been described as a method ideal for the detection of low-level mutations by increasing the sensitivity of PCR-based assays, the application of COLD-PCR for the detection of the BRAF V600E mutation could not improve the sensitivity of our assay (unpublished data). Similar results have been shown by Fadhil and colleagues.4 In addition, the intensive optimization and increased time necessary for each run (approximately 6 hours) renders COLD-PCR a sophisticated but labor-intensive method. Although the optimization of COLD-PCR is still being pursued in our laboratory, the findings thus far are rather modest. We request that the authors comment on (1) the different melting behavior of the MCF-7 wild-type DNA and the SKMel28 DNA in the difference plots and (2) the failure to refer to other studies,2 including their own work,3 showing a detection limit of 5% mutated BRAF alleles using HRM in combination with conventional PCR.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    7
    References
    4
    Citations
    NaN
    KQI
    []