Dual-Energy and Iterative Metal Artifact Reduction for Reducing Artifacts Due to Metallic Hardware: A Loosening Hip Phantom Study

2019 
OBJECTIVE. The purpose of this study was to compare the value of iterative metal artifact reduction (IMAR) with that of dual-energy CT (DECT) and filtered back projection (FBP) CT protocols for reducing metal artifacts and for facilitating visualization of the acetabular cortex in a loosening hip phantom model. MATERIALS AND METHODS. CT scans were obtained with conventional FBP and dual-source CT for two types of hip phantom. For the quantitative study, attenuation was measured by placement of ROIs in the phantoms around the metallic hardware. The differences between mean attenuation in each ROI and the actual attenuation were compared among the three CT protocols. For the qualitative study, the visibility of the acetabular cortex in the artificial loosening area of the total hip arthroplasty model and in the joint space of the bipolar hemiarthroplasty model was evaluated by measurement of the obscured cortical angle. RESULTS. In the quantitative study, attenuation differences in the bipolar hemiarthropla...