Adverse Technical Events in Home Hemodialysis

2015 
Background There is a growing interest in home hemodialysis because of its clinical benefits. However, given that patients are responsible for performing a complex medical procedure at home, adverse-event reporting is important to ensure patient safety. The purpose of this study was to describe adverse technical events in a large cohort of home hemodialysis patients. Study Design Retrospective cohort study. Setting & Participants All consecutive patients undergoing home hemodialysis at a large tertiary-care center from 1999 through 2011 (last follow-up, July 2012). Outcomes Overall rate of adverse technical events and number/rate of severe adverse events (defined as those requiring intervention). Results The cohort consisted of 202 patients with total follow-up of 757 patient-years. The cohort underwent a median of 5 dialysis treatments per week and 8 hours per session. 22 first adverse events and 7 recurrent events were identified. Adverse event rates were 0.049 per arteriovenous fistula access-year, 0.015 per arteriovenous graft access-year, and 0.022 per dialysis catheter access-year. Event rates per 1,000 dialysis treatments were 0.208, 0.068, and 0.087 for arteriovenous fistula, arteriovenous graft, and dialysis catheter access, respectively. Most adverse events were related to needle dislodgement (n=18) or air embolism (n=6). 8 adverse events resulted in emergency department visits and 5 required hospital admission. The rate of severe adverse events was 0.009 per patient-year of home hemodialysis and 0.038 per 1,000 dialysis treatments. Interventions included 3 blood transfusions, 2 catheter changes, 1 use of intravenous fluids, and 1 need for urgent dialysis. Attempts were made to retrain or review the technique in all patients with a first adverse event. Limitations Events that were not severe may have been under-reported by patients. Conclusions Serious adverse technical events in home hemodialysis are relatively rare. Strategies to further prevent these events may include patient retraining and periodic vascular access technique audit.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    17
    References
    35
    Citations
    NaN
    KQI
    []