Urine sediment examination: a comparison of automated urinalysis systems and manual microscopy.

2007 
Abstract Background Several automated instruments examining urine sediment have been introduced. We compared the performance of Sysmex UF-100 and Iris iQ200 with manual microscopy in urine sediment testing. Methods Four hundred and thirty-six urine samples were collected. The urine sediments were examined by manual microscopy and these 2 automated urinalysis systems. Results The within-run CVs for urine samples ranged from 3.4% to 22.3% for the iQ200, 1.6% to 24.2% for the UF-100 and 12.5% to 43.9% for manual microscopy. Between-run CVs on quality-control samples ranged from 6.1% to 32.4% for the iQ200 and 3.5% to 24.7% for the UF-100. The agreement between methods was good for red blood cells and white blood cells counts based on r values of 0.935 to 0.968. However, for epithelial cells, the values measured by different systems were poorly correlated ( r  = 0.888–0.922). The Bland–Altman plot indicated a trend towards the automated cell count being greater than the manual microscopy as the epithelial cell count increased. Casts were difficultly differentiated by 2 automated systems. Conclusions These 2 automated urinalysis systems demonstrated good concordance with each other in urine sediment examination. The automated process could be used as a screening procedure but some manual microscopy was still necessary.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    22
    References
    76
    Citations
    NaN
    KQI
    []