Analyzing 3-tesla magnetic resonance imaging units for implementation in radiosurgery

2005 
Object. The limiting factor affecting accuracy during gamma knife surgery is image quality. The new generation of magnetic resonance (MR) imaging units with field strength up to 3 teslas promise superior image quality for anatomical resolution and contrast. There are, however, questions about chemical shifts or susceptibility effects, which are the subject of this paper. Methods. The 3-tesla MR imaging unit (Siemens Trio) was analyzed and compared with a 1-tesla unit (Siemens Magnetom Expert) and to a 1.5-tesla unit (Philips Gyroscan). Evaluation of the magnitude of error was performed within transverse slices in two orientations (axial/coronal) by using a cylindrical phantom with an embedded grid. Deviations were determined for 21 targets in a slab phantom with known geometrical positions within the stereotactic frame. Distortions caused by chemical shift and/or susceptibility effects were analyzed in a head phantom. Inhouse software was used for data analyses. The mean deviation was less than 0.3 mm in ...