THE INFLUENCE OF THE SHAPE OF ACETABULUM AND OPTIMAL REGISTRATION POINT FOR CT-BASED NAVIGATION SYSTEM

2018 
BackgroundCT-based navigation system in total hip arthroplasty(THA) is widely used to achieve accurate implant placement. The purpose of this study was to evaluate the influence of initial error correction according to the differences in the shape of the acetabulum, and correction accuracy associated with operation approach after localization of registration points at anterior or posterior area of the acetabulum.MethodsWe set the anterior pelvic plane(APP) as the reference plane, and defined the coordinates as follows: X-axis for external direction, Y-axis for anterior direction, and Z-axis for proximal direction. APP is defined by the anterior superior iliac spines and anterior border of the pubic symphysis. We made a bone model of bilateral acetabular dysplasia of the hip, after rotational acetabulum osteotomy(RAO) on one side, and performed registration using infrared-reflective markers. At first, we registered the initial error on navigation system, and calculated the accuracy of the error correction ...