Vertebral body split fracture after a single-level cervical total disc replacement

2012 
Cervical total disc replacement (TDR) is a viable option for the surgical treatment of degenerative disc disease. This 67-year-old nonsmoking male patient underwent single-level ProDisc-C cervical TDR at C5–6 without any intraoperative problem. His radicular pain improved and he had no neck pain immediately after the operation. However, on postoperative Day 3, a radiograph demonstrated a vertical split fracture of the C-5 vertebra. This fracture was managed conservatively, and 2 years postoperatively a follow-up CT scan demonstrated stable device position and fusion of the fracture. Although the linear fracture caused no neurological symptoms or device migration, the authors advocate prudence in selection and installation of keel-design prostheses, even in a single-level cervical TDR scenario.