Reduction of radiation exposure during atrial fibrillation ablation using a novel fluoroscopy image integrated 3-dimensional electroanatomic mapping system: A prospective, randomized, single-blind, and controlled study

2015 
Objective We explored whether the use of a novel fluoroscopy image integrated 3-dimensional electroanatomic mapping (F-EAM) system could result in a reduction of overall fluoroscopy time and radiation doses during the whole procedure of atrial fibrillation (AF) ablation. Methods Eighty patients (44 men (55%); mean age 63 ± 10 years) who underwent catheter ablation due to paroxysmal AF were recruited consecutively in the present study. Patients were randomized (1:1) into 2 arms for AF ablation: one using a conventional 3-dimensional electroanatomical mapping (EAM) system and the other using the F-EAM system. Results Fluoroscopy time (10:42 [interquartile range {IQR} 8:45–12:46] minutes:seconds vs 1:45 [IQR 1:05–2:22] minutes:seconds; P 2 vs 652 [IQR 326–1489] cGy·cm 2 ; P P = .362). During follow-up (5.9 ± 1.3 months), 61 patients (76.3%) had no recurrence of atrial arrhythmias. The recurrence rate between the 2 groups did not differ. Conclusion AF catheter ablation using the F-EAM system was safe and resulted in a significant reduction of radiation exposure to patients and staff without complicating the workflow of the procedure. A near-zero fluoroscopic catheter ablation procedure could be performed without compromising acute/mid-term efficacy and safety.