Detection of ingested nitromethane and reliable creatinine assessment using multiple common analytical methods

2018 
AbstractAim: Nitromethane, found in fuels used for short distance racing, model cars, and model airplanes, produces a falsely elevated serum creatinine with standard creatinine analysis via the Jaffe method. Erroneous creatinine elevation often triggers extensive testing, leads to inaccurate diagnoses, and delayed or inappropriate medical interventions. Multiple reports in the literature identify “enzymatic assays“ as an alternative method to detect the true value of creatinine, but this ambiguity does not help providers translate what type of enzymatic assay testing can be done in real time to determine if there is indeed false elevation.Methods: We report seven cases of ingested nitromethane where creatinine was determined via Beckman Coulter® analyser using the Jaffe method, Vitros® analyser, or i-Stat® point-of-care testing. Nitromethane was detected and semi-quantified using a common clinical toxic alcohol analysis method, and quantified by headspace-gas chromatography-mass spectrometry.Results: When...
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    10
    References
    1
    Citations
    NaN
    KQI
    []