The 3DBiopsy Prostate Biopsy System: Preclinical Investigation of a Needle, Actuator, and Specimen Collection Device Allowing Sampling of Individualized Prostate Lengths Between 20 and 60 mm
2017
Objective To increase the likelihood of detecting anterior cancers within the prostate and provide a specimen that spans the length of the gland. Newly designed 17- and 15-gauge (G) biopsy needles, a variable actuator, and an integrated pathology system intended for the longer cores were developed and tested for this purpose. Materials and Methods Testing was performed comparing 2 common cannula tip grinds, a Vet-point (sharp tip) and a Menghini-point (atraumatic tip), and were tested against 18-G Bard Monopty in porcine kidney. A variable actuator was developed to fire the needle 20-60 mm and tested in cadaver prostates. Results The aggregate firings for 3 different shot lengths comparing the Vet- with the Menghini-tip cannulas demonstrated 91% vs 85.2% fill (length of specimen/length of core bed, P = .007). A 15-G trocar needle with the Vet-tip cannula also had the best performance, with an aggregate standard deviation of 6.4% across 3 firing ranges and a minimum to maximum specimen length of 81%-105% of potential fill. Cadaver testing with the Vet-tip needles in the actuator for the transrectal (17-G) and transperineal (15-G) biopsies demonstrated mean fills of 93.3% and 76.5%, respectively. The new transrectal ultrasound needle obtained a 2-fold increase in specimen length over the standard Bard device ( P Conclusion Longer and consistent cores were obtained using the new biopsy needles. Combined with an adjustable actuator, the physician can obtain specimens that include peripheral and anterior zone tissue in 1 core. Determination of cancer location on the longer specimens could enhance focal therapy planning.
Keywords:
- Correction
- Source
- Cite
- Save
- Machine Reading By IdeaReader
26
References
2
Citations
NaN
KQI