Biocompatibility of surface treated pure titanium and titanium alloy byin vivo andin vitro test

2003 
In the present study, commercial pure Ti and Ti-6Al-4V alloy specimens with and without alkali and heat treatments were implanted in the abdominal connective tissue of mice. Conventional stainless steel 316L was also implanted for comparison. After three months, their biocompatibility was evaluated byin vitro andin vivo experiments. Surface structural changes of specimens due to the alkali treatment and soaking in Hank’s solution were analyzed by XRD, SEM, XPS and AES. An apatite layer, which accelerates the connection with bone, was formed more easily on the alkali treated specimens than the non-treated specimens. The number of macrophages, which is known to increase as the inflammatory reaction proceeds, was much lower for the alkali and heat treated specimens than for the others. The average thickness of the fibrous capsule formed around the implant was much thinner for the alkali and heat treated specimens than for the others.