Impact of Interventricular membranous septum length on pacemaker need with different Transcatheter aortic valve implantation systems.

2021 
Abstract Background The need for new permanent pacemaker implantation (PPI) after Transcatheter Aortic Valve Implantation (TAVI) remains a critical issue. Membranous Septum (MS) length is associated with PPI after TAVI. The aim of this study was to identify different MS thresholds for the contemporary THV-platforms. Methods This retrospective, case-control study enrolled all patients who underwent a successful TAVI procedure with contemporary THV-platforms in the Erasmus University Medical Center between January 2016 and March 2020. The follow-up period for new PPI was 30 days. MS-length was determined by Computed Tomography. Results The study consisted 653 TAVI patients with median age 80.6 years (IQR 74.7–84.8). New PPI occurred in 120 patients (18.4%). Patients with new PPI had a shorter MS-length (2.9 mm (IQR 2.3–4.3) vs. 4.2 mm (IQR 2.9–5.7), p   6 mm a low-risk phenotype with a 6.3% PPI-rate (reference). For the Lotus valve, there was no significant difference in PPI-rates between the high-risk (45.8%, OR 3.5 [95%CI 0.8–15.1]) and low-risk group (20%). By multivariate analysis MS-length, Agatston-score, use of Lotus valve, and ECG with first-degree AV block, RBBB or bifascular block were independent predictors for new PPI. Conclusion MS-length was an independent predictor for new PPI post-TAVI. Three phenotypes were found based on MS-length. MS  30%). MS > 6 mm represented a low-risk phenotype with PPI-rate