Fundamental research on a cerenkov radiation sensor based on optical glass for detecting beta-rays

2015 
In this study, a Cerenkov radiation sensor for detecting low-energy beta-particles was fabricated using various Cerenkov radiators such as an aerogel and CaF2-, SiO2-, and Al2O3-based optical glasses. Because the Cerenkov threshold energy (CTE) is determined by the refractive index of the Cerenkov radiator, the intensity of Cerenkov radiation varies according to the refractive indices of the Cerenkov radiators. Therefore, we measured the intensities of Cerenkov radiation induced by beta-particles generated from a radioactive isotope as a function of the refractive indices of the Cerenkov radiators. Also, the electron fluxes were calculated for various Cerenkov radiators by using a Monte Carlo N-Particle extended transport code (MCNPX) to determine the relationship between the intensities of the Cerenkov radiation and the electron fluxes.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    7
    References
    0
    Citations
    NaN
    KQI
    []