MID-TERM OUTCOMES OF A 32 MM CERAMIC HEAD WITH A TITANIUM-ALLOY SLEEVE IN PRIMARY TOTAL HIP ARTHROPLASTY

2018 
IntroductionModern ceramic-on-ceramic bearings have become attractive alternatives to conventional polyethylene due to their low wear and minimal particle production. However, ceramic-on-ceramic implants have been associated with ceramic fracture and squeaking. To address these issues, large ceramic heads with a titanium-alloy sleeve have been introduced although limited data are available on their clinical outcomes. The purpose of this study was to report the midterm results of primary total hip arthroplasty using a 32mm with a titanium-alloy sleeve.Materials & MethodsWe reviewed 245 patients who had 274 total hip arthroplasties with a 32 mm ceramic head with a titanium-alloy sleeve and had been followed for more than 5 years (average, 6.5 years; range, 5–9 years). The mean patient age at the time of surgery was 55.1 years (range, 16–82 years). All operations were performed at a single center. All of the ceramic implants were hot isostatic pressed, laser-marked, proof-tested third-generation alumina (Fig...