Analytical Performance of the Sysmex HISCL HBsAg Assay and Comparison with the Roche Elecsys HBsAg II Quant Assay in the Quantification of Hepatitis B Surface Antigen

2021 
Background and Objectives: This study aims to estimate the analytical performance of the Sysmex HISCL HBsAg assay and to assess the analytical correlation with the Roche Elecsys HBsAg II quant assay with clinical samples and the WHO International Standard (IS). Materials and Methods: The intra-assay precision, linearity, assay limitation, accuracy, and comparative evaluation of the HISCL HBsAg assay were estimated. Results: Extrapolating from the plot of the average total allowable error versus the reference value, an accuracy goal of 20% would be achieved around a limit of quantification (LoQ) of 0.014867 IU/mL. The percentage of biases for each level of the WHO IS measured by the two assays were less than 15%, except for the WHO 3rd IS, for which the HISCL HBsAg assay achieved a percentage of bias of 33%. In the comparative evaluation, Passing–Bablok regression analysis did not reveal any significant deviation from linearity between the two assays (y = −48.6998 + 1.9206x; p = 0.79 by the CUSUM test for linearity). The mean difference of the quantitative HBsAg level between the two assays was 1762.5 IU/mL in the Bland–Altman plot. Conclusions: The HISCL HBsAg assay, with a highly sensitive LoQ of 0.03 IU/mL, showed similar analytical performance in HBsAg quantification to the Elecsys HBsAg II quant assay and may be helpful in obtaining better diagnoses and therapeutic strategies for treating HBV infections.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    34
    References
    0
    Citations
    NaN
    KQI
    []