Evaluation of changes in Bone Scan Index at different acquisition time‐points in bone scintigraphy

2018 
Bone Scan Index (BSI) is a validated imaging biomarker to objectively assess tumour burden in bone in patients with prostate cancer, and can be used to monitor treatment response. It is not known if BSI is significantly altered when images are acquired at a time difference of 1 h. The aim of this study was to investigate if automatic calculation of BSI is affected when images are acquired 1 hour apart, after approximately 3 and 4 h. We prospectively studied patients with prostate cancer who were referred for bone scintigraphy according to clinical routine. The patients performed a whole-body bone scan at approximately 3 h after injection of radiolabelled bisphosphonate and a second 1 h after the first. BSI values for each bone scintigraphy were obtained using EXINI boneBSI software. A total of 25 patients were included. Median BSI for the first acquisition was 0·05 (range 0-11·93) and for the second acquisition 0·21 (range 0-13·06). There was a statistically significant increase in BSI at the second image acquisition compared to the first (P 0 (54%), a clinically significant increase (>0·3) was observed. The time between injection and scanning should be fixed when changes in BSI are important, for example when monitoring therapeutic efficacy. (Less)