A controlled comparison of the BacT/ALERT® 3D and VIRTUO™ microbial detection systems

2017 
The performance of the next-generation BacT/ALERT® VIRTUO™ Microbial Detection System (VIRTUO™, bioMerieux Inc., Hazelwood, MO) was compared to the BacT/ALERT® 3D Microbial Detection System (3D, bioMerieux Inc., Durham, NC) using BacT/ALERT® FA Plus (FA Plus), BacT/ALERT® PF Plus (PF Plus), BacT/ALERT® FN Plus (FN Plus), BacT/ALERT® Standard Aerobic (SA), and BacT/ALERT® Standard Anaerobic (SN) blood culture bottles (bioMerieux Inc., Durham, NC). A seeded limit of detection (LoD) study was performed for each bottle type in both systems. The LoD studies demonstrated that both systems were capable of detecting organisms at nearly identical levels [<10 colony-forming units (CFU) per bottle], with no significant difference. Following LoD determination, a seeded study was performed to compare the time to detection (TTD) between the systems using a panel of clinically relevant microorganisms inoculated at or near the LoD with 0, 4, or 10 mL of healthy human blood. VIRTUO™ exhibited a faster TTD by an average of 3.5 h, as well as demonstrated a significantly improved detection rate of 99.9% compared to 98.8% with 3D (p-value <0.05).