Quantitative performance of 124 I PET/MR of neck lesions in thyroid cancer patients using 124 I PET/CT as reference

2018 
Background In patients with differentiated thyroid cancer (DTC), serial 124I PET/CT imaging is, for instance, used to assess the absorbed (radiation) dose to lesions. Frequently, the lesions are located in the neck and they are close to or surrounded by different tissue types. In contrast to PET/CT, MR-based attenuation correction in PET/MR may be therefore challenging in the neck region. The aim of this retrospective study was to assess the quantitative performance of 124I PET/MRI of neck lesions by comparing the MR-based and CT-based 124I activity concentrations (ACs). Sixteen DTC patients underwent PET/CT scans at 24 and 120 h after administration of about 25 MBq 124I. Approximately 1 h before or after PET/CT examination, each patient additionally received a 24-h PET/MR scan and sometimes a 120-h PET/MR scan. PET images were reconstructed using the respective attenuation correction approach. Appropriate reconstruction parameters and corrections were used to harmonize the reconstructed PET images to provide, for instance, similar spatial resolution. For each lesion, two types of ACs were ascertained: the maximum AC (max-AC) and an average AC (avg-AC). The avg-AC is the average activity concentration obtained within a spherical volume of interest with a diameter of 7 mm, equaling the PET scanner resolution. For each type of AC, the percentage AC difference between MR-based and CT-based ACs was determined and Lin’s concordance correlation analysis was applied. Quantitative performance was considered acceptable if the standard deviation was ± 25% (precision), and the mean value was within ± 10% (accuracy).
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    16
    References
    4
    Citations
    NaN
    KQI
    []