Transcatheter aortic valve implantation in patients with a small aortic annulus: performance of supra-, intra- and infra-annular transcatheter heart valves

2021 
A small aortic annulus is associated with increased risk of prosthesis–patient mismatch (PPM) after transcatheter aortic valve implantation (TAVI). Whether specific transcatheter heart valve (THV) designs yield superior hemodynamic performance in these small anatomies remains unclear. Data from 8411 consecutive patients treated with TAVI from May 2012 to April 2019 at four German centers were retrospectively evaluated. A small aortic annulus was defined as multidetector computed tomography-derived annulus area   mild was more frequent after TAVI with self-expanding THV (p = 0.04). In this large contemporary multicenter patient population, a substantial number of patients with a small aortic anatomy were left with PPM after TAVI. Self-expanding supra- and intra-annular THV demonstrated superior hemodynamics in these patients at risk, however at the cost of higher rates of residual paravalvular regurgitation.
    • Correction