Repeatability of a reference region model for analysis of murine DCE‐MRI data at 7T
2006
Purpose
To test the repeatability of a reference region (RR) model for the analysis of dynamic contrast-enhanced MRI (DCE-MRI) in a mouse model of cancer at high field.
Materials and Methods
Seven mice were injected with 106 4T1 mammary carcinoma cells and imaged eight to 10 days later on a Varian 7.0T scanner. Two DCE-MRI studies were performed for each mouse (separated by 2.5 hours). The RR model was used to analyze the data, and returned estimates on the perfusion-permeability index (Ktrans) for the RR and the tissue of interest (TOI), as well as the extravascular extracellular volume fraction (ve) for the TOI.
Results
When the first injection was compared with the second injection, all parameters tested were highly correlated (r2 = 0.90, 0.62, 0.82 for the RR Ktrans, TOI Ktrans, and TOI ve, respectively, with P < 0.001 for all). To observe a statistically significant change (at the 5% level) in a treatment study with seven animals in each group, log10 changes of 0.084 and 0.077 in the tumor Ktrans and ve, respectively, are required.
Conclusion
If a reliable arterial input function (AIF) is unavailable, the RR model is a reasonable alternative to measuring MRI contrast-agent (CA) kinetics in mouse models of cancer at high field. J. Magn. Reson. Imaging 2006. © 2006 Wiley-Liss, Inc.
Keywords:
- Correction
- Source
- Cite
- Save
- Machine Reading By IdeaReader
29
References
57
Citations
NaN
KQI