An LC-MS-MS method for the determination of indinavir, an HIV-1 protease inhibitor, in human plasma.

2001 
A method for the determination of indinavir (IDV) (L-735 524) in human plasma by LC-MS-MS is discussed, and the validation data is presented. The analyte and internal standard are isolated from plasma by a simple acetonitrile precipitation of plasma proteins followed by centrifugation. LC-tandem mass spectrometry in positive ion, multiple reaction monitoring mode used pairs of ions at m/z of 614/421 for indinavir and 628/421 for internal standard, respectively. The calibration curve had a linear range from 3.0 to 12320 ng/ml when linear least square regression weighing 1/x was applied to the concentration versus peak area plot. The advantages of this method are the fast sample preparation, wide dynamic assay range and quick analysis taking only 5 min for each sample run. The robust nature of this assay has been further verified during routine use over several months involving multiple analysts.