First 10Be measurements at Trondheim 1 MV AMS

2018 
Abstract The performance of the Trondheim 1 MV AMS system for 10 Be measurements has been investigated. While the initial acceptance tests were done using the 1 + 1+ charge state, we concluded that the 2 + 2+ charge state offers superior measurement conditions. With this setting, the source-to-detector efficiency is 5.4 ± 0.2% and a 10 Be/ 9 Be background level of 3.4 ± 0.8·10 −15 was achieved. Standard samples (n = 14) with a high 10 Be content of 10 Be/Be = 2.7·10 −11 could be reproduced to a 0.4% precision, leaving counting statistics as the limiting uncertainty for most samples. The accuracy of the system was also tested using 5 secondary standards of different 10 Be content. The results corroborate the precision achieved with the standard sample and demonstrate the accuracy of the system.