Determination of selenium using atomically imprinted polymer (AIP) and hydride generation atomic absorption spectrometry.

2013 
Abstract This paper describes selenium determination based on Se 0 preconcentration in the imprinted polymer (synthesized with 2.25 mmol SeO 2 , 4-vinylpyridine and 1-vinylimidazole) with subsequent detection on-line in HG-FAAS. During the synthesis, SeO 2 is reduced to Se (0). Therefore, there are no MIP neither IIP in the present work, thus we denominated: AIP, i.e., atomically imprinted polymers. For the optimization of analytical parameters Doehlert design was used. The method presented limit of detection and limit of quantification of 53 and 177 ng L −1 , respectively, and linear range from 0.17 up to 6 μg L −1 ( r  = 0.9936). The preconcentration factor (PF), consumptive index (CI) and concentration efficiency (CE) were 232; 0.06 mL and 58 min −1 respectively. The proposed method was successfully applied to determine Se in Brazil nuts (0.33 ± 0.03 mg kg −1 ), apricot (0.46 ± 0.02 mg kg −1 ), white bean (0.47 ± 0.03 mg kg −1 ), rice flour (0.47 ± 0.02 mg kg −1 ) and milk powder (0.22 ± 0.01 mg kg −1 ) samples. It was possible to do 12 analyzes per hour. Accuracy was checked and confirmed by analyzing certified reference material (DORM-2, dogfish muscle), and samples precision was satisfactory with RSD lower than 10%.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    55
    References
    21
    Citations
    NaN
    KQI
    []