Geometric distortion characterization and correction for the 1.0 T Australian MRI‐linac system using an inverse electromagnetic method

2020 
Purpose The magnetic resonance imaging (MRI)-Linac system combines a MRI scanner and a linear accelerator (Linac) to realize real-time localization and adaptive radiotherapy for tumors. Given that the Australian MRI-Linac system has a 30-cm diameter of spherical volume (DSV) with a shimmed homogeneity of +/- 4.05 parts per million (ppm), a gradient nonlinearity (GNL) of = 1.5-mm distortions from 6.3% to 1.3%, indicating substantially improved geometric accuracy. Conclusions The experimental results indicated that the proposed method could provide substantially improved geometric accuracy for the region outside of the DSV, when comparing with the fifth-order SH-based method.