Determination of 4-demethoxy-3′-deamino-3′-aziridinyl-4′-methylsulphonyldaunorubicin and its 13-hydroxy metabolite by direct injection of human plasma into a column-switching liquid chromatography system with mass spectrometric detection

1999 
Abstract A selective, sensitive and fully automated column-switching LC system using direct injection of human plasma followed by mass spectrometry (MS) detection was developed and validated to determine the concentrations of 4-demethoxy-3′-deamino-3′-aziridinyl-4′-methylsulphonyldaunorubicin (PNU-159548) and its 13-hydroxy metabolite (PNU-169884). A 50-μl human plasma sample was directly introduced into a C 4 -alkyl-diol silica clean-up column separating analytes from proteins and polar endogenous compounds using water and methanol as the mobile phase. The fraction containing PNU-159548 and its metabolite was back-flushed and transferred to the analytical column. The compounds were separated using a Zorbax SB C 8 column (150×4.6 mm, 5 μm) under gradient conditions with the mobile phase containing acetonitrile and 2 m M ammonium formate, pH 3.5. MS detection was by atmospheric pressure ionisation with multiple reaction monitoring in positive ion mode. Linearity was demonstrated over the calibration range of 0.051–10.291 ng/ml for PNU-159548 and 0.104–10.434 ng/ml for PNU-169884. The assay was validated with respect to accuracy, precision and analyte stability. On the basis of the validation data, the developed analytical method was found to be suitable for use in Phase I clinical studies.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    4
    References
    10
    Citations
    NaN
    KQI
    []