Image quality in coronary computed tomography angiography: influence of adaptive statistical iterative reconstruction at various radiation dose levels

2018 
BackgroundComputed tomography (CT) technology is rapidly evolving and software solution developed to optimize image quality and/or lower radiation dose.PurposeTo investigate the influence of adaptive statistical iterative reconstruction (ASIR) at different radiation doses in coronary CT angiography (CCTA) in detailed image quality.Material and MethodsA total of 160 CCTA were reconstructed as follows: 55 scans with filtered back projection (FBP) (650 mA), 51 scans (455 mA) with 30% ASIR (ASIR30), and 54 scans (295 mA) with 60% ASIR (ASIR60). For each reconstruction, subjective image quality was assessed by five independent certified cardiologists using a visual grading analysis (VGA) with five predefined image quality criteria consisting of a 5-point scale. Objective measures were contrast, noise, and contrast-to-noise ratio (CNR).ResultsThe CTDIvol resulted in 10.3 mGy, 7.4 mGy, and 4.6 mGy for FBP, ASIR30, and ASIR60, respectively. Homogeneity of the left ventricular lumen was the sole aspect in which re...