A low-cost solution for the restoration of femoral head centre during total hip arthroplasty

2013 
Restoration of joint centre during total hip arthroplasty is critical. While computer-aided navigation can improve accuracy during total hip arthroplasty, its expense makes it inaccessible to the majority of surgeons. This article evaluates the use, in the laboratory, of a calliper with a simple computer application to measure changes in femoral head centres during total hip arthroplasty. The computer application was designed using Microsoft Excel and used calliper measurements taken pre- and post-femoral head resection to predict the change in head centre in terms of offset and vertical height between the femoral head and newly inserted prosthesis. Its accuracy was assessed using a coordinate measuring machine to compare changes in preoperative and post-operative head centre when simulating stem insertion on 10 sawbone femurs. A femoral stem with a modular neck was used, which meant nine possible head centre configurations were available for each femur, giving 90 results. The results show that using this...