A novel pressure sensor with an optical system for coil embolization of intracranial aneurysms. Laboratory investigation.

2009 
Object In endovascular coil embolization for an intracranial aneurysm, the excessive pressure created during coil insertion into an aneurysm can cause a catastrophic rupture or dislodge a microcatheter tip from the aneurysm dome, resulting in insufficient embolization. Such undue mechanical pressure can only be subjectively detected by the subtle tactile feedback the surgeon experiences. Therefore, the authors of this study developed a new sensor device to measure the coil insertion pressure via an optical system. Methods This novel sensor system consists of a hemostatic valve connected to the proximal end of a microcatheter (Y-connector). The sensor principle is based on an optical system composed of a light-emitting diode (LED) and a line sensor. The latter measures how much the coil-delivery wire slightly bends in response to the insertion pressure by detecting the wire shadow. This information is translated into a given force level. Experimental aneurysm embolization was performed using this optical s...