Current of injury predicts adequate active lead fixation in permanent pacemaker/defibrillation leads.

2005 
Objectives The aim of this study was to determine whether current of injury can guide adequate placement of active-fixation pacing leads. Background Active-fixation leads cause injury to the myocardium at the time of fixation, manifested as a current of injury (COI) that may result in acute elevation of pacing thresholds. The relationship of COI to subsequent improvement in pacing thresholds is not clear. Methods Sixty-five patients undergoing active-fixation lead implantation were enrolled. Current of injury was characterized as the duration of the intracardiac electrogram (EGM) and the magnitude of ST-segment elevation. Pacing parameters were measured up to 10 min after fixation. Results A total of 96 active-fixation leads were studied, and 76 leads had a current of injury. From baseline to the time of fixation, the duration of the intracardiac EGM in ventricular leads increased from 150 ± 31 ms to 200 ± 25 ms (p Conclusions The development of a COI indicates that within 10 min of fixation, pacing threshold will return to an acceptable range even if the initial measurement is high. Conversely, without a COI, lead fixation is not adequate and the lead should be repositioned.