Optimisation des règles de déclenchement d’une revue microscopique du frottis sanguin en laboratoire de ville

2018 
: Manual blood smear review is required when an anomaly detected by the automated hematologic analyzer triggers a flag. Our will through this study is to optimize these flagging thresholds for manual slide review in order to limit workload, while insuring clinical care through no extra false-negative. Flagging causes of 4,373 samples were investigated by manual slide review, after having been run on ADVIA 2120i. A set of 6 user-adjustments is proposed. By implementing all recommendations that we made, false-positive rate falls from 81.8% to 58.6%, while PPV increases from 18.2% to 23.7%. Hence, use of such optimized thresholds enables us to maximize efficiency without altering clinical care, but each laboratory should establish its own criteria to take into consideration local distinctive features.