Dynamic T 2 mapping by multi‐spin‐echo spatiotemporal encoding
2020
PURPOSE: To develop a pulse sequence for acquiring robust, quantitative T2 relaxation maps in real time. METHODS: The pulse scheme relies on fully refocused spatiotemporally encoded multi-spin-echo trains, which provide images that are significantly less distorted than spin-echo echo planar imaging-based counterparts. This enables single-shot T2 mapping in inhomogeneity-prone regions. Another advantage of these schemes stems from their ability to interleave multiple scans in a reference-free manner, providing an option to increase sensitivity and spatial resolution with minimal motional artifacts. RESULTS: The method was implemented in preclinical and clinical scanners, where single-shot acquisitions delivered reliable T2 maps in =200 ms with approximately 250 microm and approximately 3 mm resolutions, respectively. Ca. 4 times higher spatial resolutions were achieved for the motion-compensated interleaved versions of these acquisitions, delivering T2 maps in ca. 10 s per slice. These maps were nearly indistinguishable from multi-scan relaxometric maps requiring orders-of-magnitude longer acquisitions; this was confirmed by mice head and real-time mice abdomen 7T scans performed following contrast-agent injections, as well as by 3T human brain and breast scans. CONCLUSION: This study introduced and demonstrated a new approach for acquiring rapid and quantitative T2 data, which is particularly reliable when operating at high fields and/or targeting heterogeneous organs or regions.