PURPOSE: To develop a pulse sequence for acquiring robust, quantitative T2 relaxation maps in real time. METHODS: The pulse scheme relies on fully refocused spatiotemporally encoded multi-spin-echo trains, which provide images that are significantly less distorted than spin-echo echo planar imaging-based counterparts. This enables single-shot T2 mapping in inhomogeneity-prone regions. Another advantage of these schemes stems from their ability to interleave multiple scans in a reference-free manner, providing an option to increase sensitivity and spatial resolution with minimal motional artifacts. RESULTS: The method was implemented in preclinical and clinical scanners, where single-shot acquisitions delivered reliable T2 maps in