Development of a NaI(Tl) detector with superior photon energy resolution for use above 100 MeV

1988 
Abstract We have designed and tested a new high resolution NaI(Tl) total absorption scintillation counter. The detector is a cylinder composed of a 26.7 cm diameter by 55.9 cm long NaI core with a concentric 10.8 cm thick NaI annulus that is divided into quadrants. The NaI detector is surrounded by a 12.7 cm thick plastic scintillator to veto both cosmic rays and events with significant shower leakage from the NaI. High uniformity of light production and collection throughout the detector is required for superior resolution. The detector has a measured resolution of 1.3% and 1.7% FWHM for 130 MeV photons and 330 MeV electrons, respectively. Computer simulations to account for loss of resolution due to pileup and energy spread of the beam indicate that the ultimate experimental resolutions at these energies are 1.2±0.1% and 1.3±0.1%. The resolutions at these two energies are at least a factor of 2 better than that of any other total absorption scintillation counter available today. Based on shower simulations, the detector is expected to have a resolution of approximately 1.3% for collimated 130–2000 MeV photons.