Three-dimensional true FISP for high-resolution imaging of the whole brain
2003
While high-resolution T1-weighted sequences, such as three-dimensional magnetization-prepared rapid gradient-echo imaging, are widely available, there is a lack of an equivalent fast high-resolution sequence providing T2 contrast. Using fast high-performance gradient systems we show the feasibility of three-dimensional true fast imaging with steady-state precession (FISP) to fill this gap. We applied a three-dimensional true-FISP protocol with voxel sizes down to 0.5×0.5×0.5 mm and acquisition times of approximately 8 min on a 1.5-T Sonata (Siemens, Erlangen, Germany) magnetic resonance scanner. The sequence was included into routine brain imaging protocols for patients with cerebrospinal-fluid-related intracranial pathology. Images from 20 patients and 20 healthy volunteers were evaluated by two neuroradiologists with respect to diagnostic image quality and artifacts. All true-FISP scans showed excellent imaging quality free of artifacts in patients and volunteers. They were valuable for the assessment of anatomical and pathologic aspects of the included patients. High-resolution true-FISP imaging is a valuable adjunct for the exploration and neuronavigation of intracranial pathologies especially if cerebrospinal fluid is involved.