Initial results of the commissioning of the HRIBF recoil mass spectrometer

1997 
The recoil mass spectrometer at the Holifield Radioactive Ion Beam Facility is currently undergoing commissioning tests. This new spectrometer is designed to transmit ions with rigidities of K = 100 resulting from fusion-evaporation reactions using inverse-kinematics. The device consists of two sections: a momentum separator to provide beam rejection and a mass separator for product identification. Using normal-kinematic and symmetric reactions, the commissioning tests have shown that the A/Q acceptance is almost {+-}5%, the energy acceptance is approximately {+-}12%, and there has been little, if any, primary beam observed on the focal plane. Commissioning tests are presently underway with reactions using inverse-kinematics.