Transthoracic defibrillation: Effect of sternotomy on chest impedance
1992
Abstract The purpose of this study was to determine the effect of sternotomy on transthoracic impedance, a major determinant of current flow and defibrillation success. Transthoracic impedance was determined by using a validated test-pulse technique that does not require actual shocks. Seventeen patients undergoing median sternotomy were studied prospectively. Transthoracic impedance was determined before operation, 3 to 5 days after operation and (in eight patients) ≥1 month after operation. When measured using paddle electrodes placed in the standard apex-right parasternal defibrillating position, transthoracic impedance declined after sternotomy in all patients, from 77 ± 18 to 59 ± 17 Ω (p It is concluded that transthoracic impedance declines after sternotomy. At any operator-selected energy level a higher current flow will result after sternotomy; this may facilitate postoperative defibrillation.