Clinical and radiographic outcomes with a posteriorly augmented glenoid for Walch B2, B3, and C glenoids in reverse total shoulder arthroplasty

2019 
Objective The purpose of this study is to quantify the clinical and radiographic outcomes of patients with severe posterior glenoid wear who were treated with reverse total shoulder arthroplasty (rTSA) and a posterior augmented baseplate. Methods A total of 67 primary rTSA patients with osteoarthritis and posterior glenoid wear were treated with an 8° posterior augmented glenoid baseplate. All patients had a Walch B2, B3, or C glenoid, 2-year minimum follow-up, and mean follow-up of 40 months. All patients were scored preoperatively and at the latest follow-up using 5 clinical outcome metrics; active range of motion was also measured. A Student's 2-tailed, unpaired t-test quantified differences in outcomes, where P Results All patients experienced significant improvements in pain and function after primary rTSA with a posterior augmented glenoid baseplate. Three complications were reported for a rate of 4.5%; no cases of aseptic glenoid loosening occurred. A total of 90% of patients exceeded the minimal clinically important difference threshold, and 80% of patients exceeded the substantial clinical benefit threshold for each clinical outcome metric and range of motion measure. No differences in outcomes or complications were observed between Walch B2 and B3 patients, demonstrating that this full-wedge posterior augmented baseplate was equally good in each type of glenoid deformity. Discussion Primary rTSA patients with Walch B2, B3, and C glenoids who received an 8° posterior augmented glenoid baseplate experienced excellent clinical and radiographic outcomes with a low complication rate and no reports of aseptic glenoid loosening at a mean follow-up of 40 months.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    43
    References
    18
    Citations
    NaN
    KQI
    []