language-icon Old Web
English
Sign In

High energy electron fluxmeter

1993 
An instrument for measurement of high energy electrons (1-10 MeV) in the Earth's radiation belts has been designed, calibrated and flown on-board the CRRES spacecraft. The instrument uses a triple coincidence arrangement with two solid state detectors and a BGO scintillator crystal to identify electrons and discriminate against energetic protons. A thorough description of the construction and calibration is presented along with an example of on-orbit data. >
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    4
    References
    19
    Citations
    NaN
    KQI
    []