Analysis of the specificity of the SD Biosensor Standard Q Ag-Test based on Slovak mass testing data

2020 
From 31.10. - 1.11.2020 Slovakia has used the SD Biosensor Standard Q Ag-Test for nationwide tests for SARS-CoV-2, in which 3,625,332 persons from 79 counties were tested. Based on this data, we calculate that the specificity of the test is at least 99.6% (with a 97.5% one-sided lower confidence bound). Our analysis is based on a worst case approach in which all positives are assumed to be false positives. Therefore, the actual specificity is expected to exceed 99.6%.