Factors Associated with Glycemic Control in Youth with Type 1 Diabetes Transitioning from Pediatric to Adult Care
2013
s / Can J Diabetes 37 (2013) S13eS84 S42 reference analyzer. Accuracy was assessed per ISO 15197:2003 criteria ( 95% of results within 0.8 mmol/L [15 mg/dL] or 20% of reference at glucose concentrations 99% of results were within 0.6 mmol/L (10 mg/dL) or 10% of the reference value. Precision testing showed a pooled coefficient of variance (CV) of 1.2% to 2.4% for blood glucose concentrations 2.4 to 18.2 mmol/L (43.0 to 328.4 mg/dL). Conclusion: The CONTOUR NEXT LINK BGMS, which wirelessly communicates with certain Medtronic insulin pumps and continuous glucose monitoring systems, met and exceeded both ISO 15197:2003 and ISO 15197:2013 accuracy criteria. The BGMS also demonstrated precision with a low CV.
Keywords:
- Correction
- Source
- Cite
- Save
- Machine Reading By IdeaReader
0
References
0
Citations
NaN
KQI