[Estimation of exposure dose on MDCT examination--the measurement of organ dose and effective dose by anthropomorphic phantom].

2010 
: In order to evaluate the exposure dose in CT examinations, we measured the tissue and organ doses by test site in 4-row, 16-row, and 64-row multi detector CT by using an anthropomorphic phantom and fluorescent glass dosimeters. Furthermore, we calculated the effective dose by using the tissue weighting factor recommended by the ICRP in 2007. The effective dose in the head and neck examinations was 1.4-3.1 mSv, whereas the maximum skin dose was 278.9 mGy in head perfusion CT. The effective dose in examinations of the body trunk was 10.1-35.2 mSv. In addition, the organ dose and skin dose in the scanning range was similar to the CTDI(vol) in head and neck examinations, while it was higher than the CTDI(vol) in examinations of the body trunk. The exposure dose of patients undergoing CT is high in comparison to other radiological examinations. As a result, due to consecutive examinations, an absorbed dose of more than 100 mGy is possible. A future problem therefore remains how to lower the overall exposure dose with the introduction of new radiographic diagnostic modalities, such as phase scan or coronary CT angiography.